Skip to the main content.

Medical Imaging Equipment

We carry CT, MRI, PET/CT, C-arm, O-arm, Cath lab, X-ray, Mammography, and Ultrasound from major providers like Siemens, GE, Philips, Toshiba, Neusoft, Hologic, and more.

Company

Block Imaging is the Multi-Vendor Service, Parts, and Equipment Provider that keeps your systems reliable, costs down, and you in control.

About Block Imaging
Contact Us

Finding Artifacts in Your Mammo Image? Read This.

If you’re seeing artifacts on your images, it can be alarming, especially when image quality is critical. The good news is that some artifact issues can be resolved without immediately calling for service. In many cases, a careful cleaning or a rerun of the gain calibration can correct the problem. This guide walks through the basic troubleshooting process so you can try the simplest fixes first before escalating the issue.

Need immediate Mammo assistance? Call +1 877-733-7142

Clean the Components

Start by taking a microfiber cloth and cleaning the flat field phantom and the breast platform. Make sure to remove any visible dust, debris, or residue that may be sitting on the surfaces. Even small particles can create image artifacts, so this step should be done carefully. 

Next, clean the detector surface as well. Then wipe both sides of the flat field phantom thoroughly to make sure no debris is left behind on either side. 

Clean flat field phantom and breast platform

After cleaning all of these surfaces, take another image and check to see whether the artifact is still present. If the artifact is gone, then the issue was likely caused by debris or contamination, and you’ve solved the problem without any further troubleshooting.

Before and After Artifact Clean

 

Check the Gain Calibration

If the artifact is still showing after cleaning, the next step is to consider whether it may have been introduced during the last gain calibration. In some cases, an artifact can become “gained in” or burned into the calibration process. Before rerunning the gain calibration, clean the flat field phantom and breast platform again to make sure the calibration is performed with completely clean surfaces. 

Once the surfaces are clean, go ahead and rerun the gain calibration following your system’s standard process. 

After the gain calibration is complete, test the system again by taking another image and checking for the artifact. If the artifact has disappeared, then the calibration process resolved the issue.

 

Still Having Issues?

If the artifact is still present after both a full cleaning and a rerun of the gain calibration, the issue likely requires further troubleshooting from a qualified field service engineer. At that point, contact service support to determine whether the issue can be resolved remotely or if an on-site visit is needed.


Need help? No worries: our Service Team is standing by to assist. Give us a call at (877) 733-7142 or use the button below to contact us and let us know what you need.

Contact Us

 

 

Picture of Noah Ford

Noah Ford

Noah has worked with medical imaging equipment since 2018. He specializes in Hologic, GE, and Siemens mammo systems. He has a passion for training up and equipping the next generation of Field Service Engineers through Block Imaging Apprentice programs. After work, you can find him volunteering at his local church, playing board games, and spending time with family.